美國戴安Dionex OmniPac PCX-500 Mixed-Mode Column 042195 液相色譜柱
如果您對該產品感興趣的話,可以
產品名稱: 美國戴安Dionex OmniPac PCX-500 Mixed-Mode Column 042195 液相色譜柱
產品型號: OmniPac PCX-500 042195
產品展商: 戴安Dionex
產品文檔: 無相關文檔
簡單介紹
戴安公司生產的儀器主要有離子色譜儀、高效液相色譜儀、快速溶劑萃取儀和氨基酸直接分析儀。其中離子色譜儀憑借著雄厚的技術優(yōu)勢和方法已廣泛地應用于各個領域,其市場占有率達80%。全球與離子色譜相關的文章有90%均出自戴安公司生產的離子色譜儀。
戴安公司的快速溶劑萃取儀15分鐘之內僅用15毫升溶劑既可完成對10g樣品的萃取,該技術已在環(huán)境、**、食品和聚合物工業(yè)等領域得到廣泛應用。
美國戴安Dionex OmniPac PCX-500 Mixed-Mode Column 042195 液相色譜柱
的詳細介紹
OmniPac PCX-500 Mixed-Mode Column
The OmniPac? PCX-500 column is a macroporous, high-capacity, solvent-compatible
column designed to separate cationic and neutral species in a single
run. This column combines cation-exchange and polymeric reversed-phase
retention mechanisms in a single column.
-
Good for the simultaneous separation of neutral and cationic species
-
Acid and base compatible, pH 0–14
-
100% solvent compatible
-
Not as effective as the CS12A for inorganic cations
The
multimode separation mechanism is achieved by coating a macroporous
hydrophobic core for reversed-phase retention with a cation-exchange
latex bead layer for cation-exchange retention. The macroporous
structure provides an accessible hydrophobic core where reversed-phase
retention occurs. The cation-exchange selectivity is provided by a
cation-exchange MicroBead? layer that coats the hydrophobic
core. Since the cation-exchange layer is identical to that of the
OmniPac PCX-100, the two columns exhibit similar ion-exchange
selectivity. As a result, the OmniPac PXC-100 can be used to develop the
ion-exchange separation, and the method can then be transferred to the
OmniPac PXC-500 for further refinement to include solvent for the
reversed-phase separation. The OmniPac PCX-500 is 100% solvent
compatible and should always be run with at least 1% organic solvent in
the mobile phase to maintain column life.
The OmniPac PCX-500 may
be used for the separation of compounds such as adrenergics, alcohols,
anilines, antidepressants, anti-inflammatory agents, plating bath
brighteners, cephalosporins, stains and dyes, herbicides, nucleotides
and nucleosides, peptides, sulfonamides, and water-soluble vitamins.
技術參數:
OmniPac PCX-500 Mixed-Mode Column Specifications
|
Dimensions
|
OmniPac PCX-500 Analytical Column:
4 × 250 mm
OmniPac PCX-500 Guard Column:
4 × 50 mm
|
Maximum operating pressure
|
4000 psi (27.9 MPa)
|
Recommended operating temperature
|
Ambient
|
Typical operating conditions
|
1.0 mL/min with at least 1% organic solvent
Recommended flow rate: 1.0 mL/min
|
Mobile phase compatibility
|
100%
compatible with common organic solvents. Recommended eluents: 1%
solvent. Acid to maintain weak bases in the cationic form, salt (e.g.,
Na+ for cation exchange).
|
Resin composition
|
8.5-μm-diameter
ethylvinylbenzene/divinylbenzene macroporous substrate (55%
cross-linked) agglomerated with 200-nm MicroBead sulfonic
acid-functionalized latex (5% cross-linked).
|
Anion-exchange capacity
|
120 μeq/column (4 × 250 mm analytical column)
|
Column construction
|
PEEK with 10-32 threaded ferrule-style end fittings. All components are nonmetallic.
|
訂購信息:
Analytical Columns
|
OmniPac PCX-500 Analytical Column (4 x 250 mm)
|
042191
|
Guard Columns
|
OmniPac PCX-500 Guard Column (4 x 50 mm)
|
042195
|