美國戴安Dionex IonPac AS20 Column 075399陰離子交換色譜柱
如果您對該產(chǎn)品感興趣的話,可以
產(chǎn)品名稱: 美國戴安Dionex IonPac AS20 Column 075399陰離子交換色譜柱
產(chǎn)品型號(hào): 075399
產(chǎn)品展商: 戴安Dionex
產(chǎn)品文檔: 無相關(guān)文檔
簡單介紹
戴安公司生產(chǎn)的儀器主要有離子色譜儀、高效液相色譜儀、快速溶劑萃取儀和氨基酸直接分析儀。其中離子色譜儀憑借著雄厚的技術(shù)優(yōu)勢和方法已廣泛地應(yīng)用于各個(gè)領(lǐng)域,其市場占有率達(dá)80%。全球與離子色譜相關(guān)的文章有90%均出自戴安公司生產(chǎn)的離子色譜儀。
戴安公司的快速溶劑萃取儀15分鐘之內(nèi)僅用15毫升溶劑既可完成對10g樣品的萃取,該技術(shù)已在環(huán)境、**、食品和聚合物工業(yè)等領(lǐng)域得到廣泛應(yīng)用。
美國戴安Dionex IonPac AS20 Column 075399陰離子交換色譜柱
的詳細(xì)介紹
IonPac AS20 Hydroxide-Selective Anion-Exchange Column
For Determination of Trace Perchlorate Using Suppressed Conductivity Detection
The IonPac? AS20 hydroxide-selective anion-exchange column is specifically designed
for determination of trace concentrations of perchlorate in drinking
water, surface water, and groundwater matrices. The capacity and
selectivity of the AS20 ensures that perchlorate can be quantified at
low μg/L concentrations using suppressed conductivity detection, even in
the presence of very high concentrations of chloride, carbonate, and
sulfate. The new IonPac AS20 Capillary Column is packed with the same
material as the equivalent standard bore version (providing the same
performance as a 4 mm column) but requires only 1/100th the eluent flow
rate. The capillary format provides the advantage of significantly
reduced eluent consumption, which helps reduce operating costs.
-
Recommended for US EPA Method 314.1 (Confirmatory Method)
-
High capacity: 310 μeq per column. (4 × 250 mm)
-
Use the Cryptand C1 Concentrator Column for sample preconcentration with the analytical column
-
Simplified Reagent-Free IC operation provided by the Eluent Generator
-
ASRS 300 or ACES 300 provides RFIC operation with low background and enhanced analyte sensitivity
-
Column selectivity is optimized for 30 °C operating temperature
-
Compatible with organic solvents
The
IonPac AS20 is the specified column in US EPA Method 314.1
(Confirmatory Method). The Cryptand C1 Concentrator Column is the
specified concentrator column for sample preconcentration in that
method.
Use the IonPac AS20 column with an RFIC? system for optimum detection limits. Use the Anion Self-Regenerating Suppressor (ASRS? 300) or the Anion Capillary Electrolytic Suppressor (ACES 300) with the AS20 column for eluent suppression.
IonPac AS20 Anion-Exchange Column Specifications
|
Maximum Operating Pressure:
|
4000 psi (standard and microbore) 5000 psi (capillary)
|
Mobile Phase Compatibility:
|
pH 0–14; 0–100% HPLC solvents
|
Ion Exchange Group:
|
Functional Group: Alkanol quaternary ammonium ion
|
Column Construction:
|
PEEK with 10–32 threaded ferrule style end fittings.
All components are nonmetallic.
|
Functional Group Characteristics:
|
Ultralow Hydrophobicity
|
Dimensions:
|
IonPac AS20 Capillary Column:
0.4 ×250 mm
IonPac AG20 Capillary Guard Column:
0.4 × 50 mm
IonPac AS20 Analytical Column:
2 × 250 mm
IonPac AG20 Guard Column:
2 × 50 mm
IonPac AS20 Analytical Column:
4 × 250 mm
IonPac AG20 Guard Column:
4 × 50 mm
|
Capacity:
|
3.1 μeq (0.4 × 250 mm capillary column)
77.5 μeq (2 × 250 mm analytical column)
310 μeq (4 × 250 mm analytical column)
0.06 μeq*(0.4 × 50 mm capillary guard column)
1.5 μeq* (2 × 50 mm guard column)
6 μeq* (4 × 50 mm guard column)
*Guards are packed with a low-capacity microporous resin.
|
Substrate Characteristics:
|
Bead Diameter and Pore Size:
7.5 μm (2 × 250, 4 × 250 and 0.4 × 250 mm);
super macroporous resin (2000 ?)
11 μm (2 × 50, 4 × 50 and 0.4 × 50 mm);
microporous resin (<1 ?)
Crosslinking (%DVB): 55%
|
Analytical Columns
|
IonPac AS20 Analytical Column (2 x 250 mm)
|
063065
|
IonPac AS20 Analytical Column (4 x 250 mm)
|
063148
|
IonPac AS20 Capillary Column (0.4 x 250 mm)
|
075399
|
Guard Columns
|
IonPac AG20 Guard Column (2 x 50 mm)
|
063066
|
IonPac AG20 Guard Column (4 x 50 mm)
|
063154
|
IonPac AG20 Capillary Guard Column (0.4 x 50 mm)
|
075400
|